Accuracy will largely depend on calibration standards used. A good set of those that goes up to 6 GHz will cost many times the cost of this device.
by avian 4y ago
Accuracy will largely depend on calibration standards used. A good set of those that goes up to 6 GHz will cost many times the cost of this device.
- maidchan375 4y agoThis is true. The VNA can only provide repeatability and precision, and it's the calibration kit's job to provide accuracy.
- MegaDeKay 4y agoThat's part of it, but linearity and the device's own return loss (a good cal kit can only do so much) are also substantial factors.
- avian 4y agoIs linearity a problem in these sorts of devices? I'm not sure where it would come into play. I'm assuming this does IF processing in software and it doesn't have a classical detector where linearity would be a problem. Yes, return loss and directivity of the bridges/directional couplers in there is a big factor. It directly corresponds to usable dynamic range after calibration. If your directivity is only 20 dB at 6 GHz, 110 dB of dynamic range at the detector won't do anything for the quality of the final measurement. So an honest way of specifying dynamic range would be for a fully calibrated device. Giving that they say it has 110 dB, I don't think that's the case here.
- maidchan375 4y agoThe linearity is shown in the "Example measurements" graphs