Indeed, there is no way to remove the physical stop. Furthermore, the design process is shaped so that it leads to such designs. If the first iteration of the
by notalaser 10y ago
Indeed, there is no way to remove the physical stop.
Furthermore, the design process is shaped so that it leads to such designs. If the first iteration of the design had included a removable physical stop, it would have been caught in the risk analysis. Most mission-critical fields have standards that specify this sort of stuff (IEC 60601 for medical devices, DO-254 and DO-178C for avionics etc.)
When designing this kind of devices, whether it's the fault of the device or not is not too relevant in general. You have to mitigate any kind of unacceptable risk (i.e. things that lead to injury or death).
There are certain common-sense exceptions here. For instance, the device isn't expected to operate properly outside its specified operation conditions, but you have to clearly state what those conditions are (altitude, temperature, humidity, presence or absence of liquids etc.) and put warning labels regarding their breach in the manual). Similarly, the level of mitigation is often specified by standards. E.g. for medical devices, IEC 60601 specifies insulation requirements that will protect against the kind of shocks you could get from a faulty network, but not if the device is struck by lightning or strapped to an electric chair. IEC 61010 (Safety requirements for electrical equipment for measurement, control and laboratory use) similarly includes provisions for the kind of protection you would need in equipment that falls under a specific type of use (e.g. here on Earth, not up there in space).